[email protected]
+44 20 3371 7320 Mon - Sat: 9:00 AM - 6:00 PM
Home All Products Spectrophotometer Double Beam UV/Vis Spectrophotometer LD-LUS-B16
Double Beam UV/Vis Spectrophotometer LD-LUS-B16

Double Beam UV/Vis Spectrophotometer LD-LUS-B16

SKU: LD-LUS-B16

Double beam UV/Vis spectrophotometerLD-LUS-B16is fully automated with improved and optimized optical path design to give highly accurate and reliable measurements. This double beam spectrophotometer offers four customized spectral bandwidth ranges from (4 nm, 2 nm, 1 nm, 0.5 nm) to meet the market needs. The wide range of measurement methods e.g. wavelength scanning, time scanning, microarray measurement meets the different range of applications.

  • ✓ Wide wavelength range of 190 - 1100 nm
  • ✓ 4 spectral bandwidth (4.0 nm, 2.0 nm, 1.0 nm, 0.5 nm)
  • ✓ 6 inches large LCD screen to display measurement parameters & readings
  • ✓ Improved and optimized optical design
  • ✓ Automatic wavelength adjustment
  • ✓ Wide range of measurement methods: wavelength scanning, time scanning, microarray measurement
  • ✓ Adapted with PC software to measure more precise measurements
  • ✓ Large scale integrated circuit design to improve the scalability and reliability of the system
  • ✓ Measurement data can be output through the printer with USB interface
  • ✓ Improved scalability and reliability of the system through LSI design
Also available in:
Wavelength Range : 190 ~ 1100 nm Wavelength Range : 190 ~ 1100 nm Wavelength Range : 190 ~ 1100 nm Wavelength Range : 190 ~ 1100 nm Wavelength range : 190 ~ 1100 nm Wavelength range : 190 ~ 1100 nm
Noise± 0.0002A @500nm
Power120 W
Weight28 kg
Dimensions560 × 450 × 230 mm
Display LCD6 inches blue LCD
Stray Light≤ 0.03 % τ (220 nm NaI, 340 nm NaNO2)
Optical SystemDouble beam
Photometric ModeT,A,C,E
Wavelength Range190 ~ 1100 nm
Baseline Flatness± 0.001 A
Photometric Range(-4) ~ 4 A
Power RequirementAC 220 V / 50 Hz, 110 V / 60 Hz.
Setting WavelengthAutomatic
Spectral Bandwidth0.5 / 1.0 / 2.0 / 4.0 nm adjustable
Wavelength Accuracy± 0.3 nm
Photometric Accuracy± 0.3 % τ(0 - 100 % τ) ; ± 0.002 A (0 ~ 0.5 A) ; ± 0.003 A(0.5 A ~ 1 A)
Wavelength Repeatability≤ 0.1 nm
Photometric Repeatability0.15 % τ(0 -100 % τ); ± 0.001 A(0 ~ 0.5 A) ; ± 0.0015 A(0.5 A ~ 1 A)

The double beam UV/VIS spectrophotometer has applications in pharmaceutical analysis, biological research, chemistry, environmental science, food science, medicine & other fields.

Similar Products