Analytical-Instruments > Spectrophotometer > Visible-Spectrophotometer > Nano Spectrophotometer LD-LNS-A10
Features
- 0.5 µl - 2.0 µl micro-volume sample for direct measurement onto pedestal
- Wavelength range 200 - 800 nm with scan capability within 5 seconds
- User friendly software
- Long life Xenon flash lamp (10 flashes up to 10 years)
- Highly concentrated samples can be used
- Direct Concentration display
- No cell or cuvettes are required
Nano Spectrophotometer LD-LNS-A10
Nano spectrophotometer with full spectrum light covers wide wavelength range of 200 to 800nm. It requires only 0.5 - 2.0 µl micro-volume sample for measurement which takes place within 5 seconds. It utilizes CCD detector, Xenon flash lamp as light source which has high stability and long operating life. They are used to detect micro volume quantities of nucleic acids and other concentrated or dilute components.
SKU Code: LD-LNS-A10
Also Available in below capacities
Wavelength range : 200 - 800 nm
Wavelength range : 200 - 850 nm
Wavelength range : 200 - 850 nm
Wavelength range : 230 nm, 260 nm, 280 nm
Wavelength range : 230 nm, 260 nm, 280 nm
Cuvette : No
- Specifications
- Product Applications
| Weight | 2.6 kg |
| Dimension | 200 x 250 x 166 mm |
| Path length | 0.2 mm (For high concentration measurement); 1.0 mm (For ordinary) |
| Light source | Xenon flash lamp |
| Detector Type | 3864-element linear silicon CCD array |
| Absorbance range | 0.02 - 90 A (10mm equivalent) |
| Measurement time | < 8 s |
| Wavelength range | 200 - 800 nm |
| Operating Voltage | 24 V DC |
| Absorbance accuracy | 1 % ( 7.332 Abs at 260 nm) |
| Minimum Sample size | 0.5 - 2.0 µl |
| Wavelength accuracy | ± 1 nm |
| Absorbance precision | 0.003 Abs |
| Wavelength Resolution | ≤ 3 nm (FWHM at Hg 546 nm) |
| Software Compatibility | Windows 7, Windows XP |
| Sample Pedestal Material | Aluminum alloy and Quartz fiber |
| Standby Power Consumption | 5 W |
| Detects nucleic acid up to | 2 - 4500 ng/µL (dsDNA) |
| Operating Power Consumption | 20 W |
| Nano spectrophotometer are used in detection of micro volume quantities of DNA/RNA and chemicals , liquids and its components in laboratory, food industries, chemistry, microbiology, medicine research and development. |