Analytical-Instruments > Spectrophotometer > UVVis-NIR-Spectrophotometer > Single Beam UV/Visible Spectrophotometer LD-LUS-A22
Features
- Wide wavelength range of 190 - 1100 nm
- 4 spectral bandwidth (4.0 nm, 2.0 nm, 1.0 nm, 0.5 nm)
- 6 inches large LCD screen to display measurement parameters & readings
- Improved and optimized optical design, light source and receiver system
- Automatic wavelength adjustment
- Wide range of measurement methods: wavelength scanning, time scanning, microarray measurement
- Adapted with PC software to measure more precise measurements
- Large scale integrated circuit design to improve the scalability and reliability of the system
- Measurement data can be output through the printer with USB interface
Single Beam UV/Visible Spectrophotometer LD-LUS-A22
Single Beam UV/Visible SpectrophotometerLD-LUS-A22comes with a wide wavelength range of 190 - 1100 nm with adjustable spectral bandwidth of 0.5 nm, 1 nm, 2 nm, 4 nm. It has a PC software to provide reliable measurements results. The wide range of measurement methods e.g. wavelength scanning, time scanning, microarray measurement meets the different range of applications.
SKU Code: LD-LUS-A22
Also Available in below capacities
Wavelength Range : 190 ~ 1100 nm
Wavelength Range : 190 ~ 1100 nm
Wavelength Range : 190 ~ 1100 nm
Wavelength Range : 190 ~ 1100 nm
Wavelength Range : 190 ~ 1100 nm
Wavelength Range : 190 ~ 1100 nm
- Specifications
- Product Applications
| Noise | ± 0.0002 A at 500 nm |
| Power | 120 W |
| Weight | 28 kg |
| Detector | Imported Silicon Photodiode |
| Stability | 0.0005 A / h at 500 nm |
| Dimensions | 560 × 450 × 230 mm |
| Display LCD | 6 inches blue LCD |
| Stray Light | ≤ 0.05 % T at 220 nm, 360 nm |
| Light Source | Imported Deuterium Lamp & Tungsten Halogen Lamp |
| Photometric Mode | T,A,C,E |
| Wavelength Range | 190 ~ 1100 nm |
| Baseline Flatness | ± 0.001 A |
| Photometric Range | (-4)A to 4 A |
| Power Requirement | AC 220 V / 50 Hz, 110 V / 60 Hz. |
| Spectral Bandwidth | 0.5 nm, 1 nm, 2 nm, 4 nm adjustable |
| Wavelength Accuracy | ± 0.3 nm |
| Photometric Accuracy | ± 0.3 % τ(0 - 100 % τ) ± 0.002 A (0 to 0.5 A) ± 0.003 A(0.5 A to 1 A) |
| Wavelength Repeatability | ≤ 0.1 nm |
| Photometric Repeatability | 0.15 % τ(0 - 100 % τ); ± 0.001 A(0 to 0.5 A); ± 0.0015 A(0.5 A to 1 A) |
| Single beam UV/VIS spectrophotometer has applications in pharmaceutical analysis, biological research, chemistry, environmental science, food science, medicine & other fields. |