ICP SpectrometerLD-LICP-B10is an instrument which measures elements in ppb levels with least interference and reduces the time taken for analysis by detecting 5 to 6 elements in 1 min. Excellent analytical precision, high optical resolution, fully automated, wide wavelength range with 2400 line grating, powerful analysis software, high performance solid-state RF system, high accuracy etc. are some of the characteristic features. The equipment consist of a Scott-type spray chamber, quartz torch, spectrophotometer with Czerny-Turner monochromator, detector with photomultiplier tube and data processing system which allows multi-element measurement along with high resolution. It finds applications in environmental, food safety, metallurgy, pharmaceutical etc. fields.
SKU Code: LD-LICP-B10
| Humidity | ≤ 70% |
| Resolution | ≤ 0.015nm (3600 line grating) ≤ 0.030nm (2400 line grating) |
| Test speed | 5 to 8 elements / min |
| Temperature | 15°C to 30°C |
| Focal length | 1000 mm |
| Output power | 800 W to 1600 W, adjustable with power efficiency more than 65% |
| Other Details | |
| Repeatability | short-term stability- RSD ≤ 1.5% Long-term stability: RSD ≤ 2% |
| Spray chamber | Scott double pass spray chamber |
| Induction coil | 25 mm × 3 ID (ID-internal diameter), equipped with three concentric quartz torch tubes 35 mm ED (ED-external diameter) |
| Control Circuit | |
| Power Stability | 220 ± 10V; 50 to 60 V |
| Wavelength range | 195 to 500 nm for 3600 line grating 195 to 800 nm for 2400 line grating |
| Signal acquisition | VF conversion |
| Working Environment | |
| Negative high voltage | -50 to -1000 V |
| Grating specifications | huge holographic grating with 3600 L/mil or 2400 L/mil and 80 mm × 100 mm of ruling area |
| Output power stability | ≤ 0.05% |
| Circuit measuring range | (10-12 to 10 -4) A |
| Technical Specifications | |
| Monochromator Specifications | |
| Solid State Power Specifications | |
| Humidity for storage and transportation | ≤ 70% |
| Limits of detection (LOD, µg/L) for most elements | 1ppb to 10ppb |
| It used for quantitative and qualitative analysis of elements in the given sample. Various fields like petrochemicals, pharmaceuticals, agriculture, environmental, food industry, metallurgy, processing industry etc. require it for several operations. |