Flammability Tester LFLC- A11 is a horizontal and vertical flame chamber with PLC touch screen control panel. The built-in thermocouple calibration and temperature control system maintains high degree of automation. It has automatic ignition and temperature function that display temperature and time. Applicable to the single wire and cable, and soft-line horizontal, vertical combustion test. Adjustable burning time and gas flow that can be automatic / manual, vertical / horizontal switching. Designed according to the relevant provisions of UL1581-2006 1080 (VW -1 combustion test) and 1100 section (horizontal sample/FT2 combustion test) standards. It is convenient to observe and reliable.
SKU Code: LD-LFLC-A11
Power | 500 W |
Burn time | 0 to 999 s (adjustable) |
Gas press | Adjustable |
Gas supply | Methane or prepared gas |
Net weight | 300 kg |
Flame angle | Adjustable |
Flame height | 20 to 110 mm/min (adjustable) |
Gross weight | 400 kg |
Power supply | AC 220 V, 50 / 60 Hz |
Gas flowmeter | 1000 ml / min |
Nozzle length | 1.65 mm |
Control system | PLC touch screen display |
Test precision | 0.01 s |
Thermal sensor | 10.0 ± 0.05 g and thermocouple electrolytic copper piece |
Nozzle diameter | 9.5 mm |
After flame time | Adjustable |
Air inlet height | 110 mm |
Cabinet capacity | 4 cubic meter with glass watch door |
Thermocouple type | K type thermocouple (with 0.5 mm diameter stainless steel pipe coating) |
Interior lamp source | LED lamp |
Thermocouple diameter | 4 mm |
Test flame rising temperature | 102°C to 103°C in 56 s |
Sample vertical moving distance | Same as test standard UL1581 |
External dimension (W × D × H) | 2100 × 1360 × 2200 mm |
Sample horizontal moving distance | Same as test standard UL1581 |
It is suitable for testing the flame retardant properties of insulation layer of the cable under abnormal condition such as overheat and overflow. |