| pH Mode | | pH Range : -2.000 to 20.000 pH | | pH Accuracy : ±0.002 pH | | pH Resolution : 0.001 / 0.01 / 0.1 pH, selectable | | pH Calibration Points : 1 to 5 points | | pH Buffer Solutions : USA / NIST / DIN / 2 custom buffers |
|
| TDS Mode | | TDS Range : 0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt | | TDS Accuracy : ±1% F.S. | | TDS Conversion Factor : 0.1 to 1.0 (default 0.5) | | TDS Resolution : 0.01, 0.1, 1, automatic |
|
| Temperature | | Temperature Measurement Range : 0 to 105 °C / 32 to 221 °F | | Temperature Measurement Accuracy : ±0.5 °C / ±0.9 °F | | Temperature Measurement Resolution : 0.1 °C/°F | | Temperature Measurement Calibration Points : 1 point (calibration range- Reading ±10 °C/°F) |
|
| Conductivity | | Conductivity Range : 0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm | | Conductivity Accuracy : ±0.5% F.S. | | Conductivity Resolution : 0.001, 0.01, 0.1, 1, automatic | | Conductivity Calibration Points : 1 to 5 points | | Conductivity Solutions : 10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS | | Conductivity Temperature Coefficient : 0.0 to 10.0% / °C | | Compensation Mode : Yes (Linear and pure water) | | Cell Constant : K=0.1, 1, 10 or Custom | | Reference Temperature : 20 or 25 °C |
|
| ORP(mV) Mode | | mV Range : -1999.9 to 1999.9 mV | | mV Accuracy : ±0.2 mV | | mV Resolution : 0.1 / 1 mV, selectable | | mV Calibration Points : 1 point | | Measurement Modes : Relative / Absolute mV |
|
| Salinity Mode | | Salinity Range : 0.00 to 80.00 ppt / 0.00 to 42.00 psu | | Salinity Resolution : 0.01 | | Salinity Accuracy : ±1% F.S. |
|
| Resistivity Mode | | Resistivity Range : 0.00 to 20.00 MΩ | | Resistivity Resolution : 0.01 | | Resistivity Accuracy : ±1% F.S. |
|
| Ion Concentration | | Ion Concentration Range : 0.001 to 19999 ppm, mg/l, mol/l, mmol/l (depending on the range of ISE) | | Ion Concentration Accuracy : ±0.5% F.S. (monovalent), ±1% F.S. (divalent) | | Ion Concentration Resolution : 0.001, 0.01, 0.1, 1, automatic | | Ion Concentration Calibration points : 2 to 5 points | | Calibration Solutions : 0.001 / 0.01 / 0.1 / 1 / 10 / 100 / 1000 / 10000 |
|
| Other Specifications | | Temperature Compensation Range : 0 to 100 °C / 32 to 212 °F manual or automatic | | Data Hold Function : Manual / Auto-read measurement endpoint | | Stability Criteria : Low / High | | Calibration Due Alarm : 1 to 31 days or off | | Output : USB communication interface | | Memory : Stores up to 500 data sets | | Connector : BNC, 6-pin mini-DIN, 3.5 mm jack socket | | Display : Custom LCD (125 × 100 mm) | | Power Supply : 5V DC | | Dimension (L × W × H) : 210 × 188 × 60 mm | | Weight : 1.5 kg |
|
| Dissolved Oxygen Mode | | Dissolved Oxygen Range : 0.00 to 20.00 mg/L | | Dissolved Oxygen Accuracy : ±0.2 mg/L / ±2.0 % | | Dissolved Oxygen Resolution : 0.01 / 0.1, selectable | | Dissolved Oxygen Calibration Points : 1 or 2 points | | % Saturation Range : 0.0 to 200.0 % saturation | | Pressure Correction : 60.0 to 112.5 kPa / 450 to 850 mmHg, manual | | Salinity Correction : 0.0 to 50.0 g/L, manual |
|