[email protected]
+44 20 3371 7320 Mon - Sat: 9:00 AM - 6:00 PM
Home All Products Water Testing Benchtop Multi-parameter Water Quality Meter LD-LMPWM-A42
Benchtop Multi-parameter Water Quality Meter LD-LMPWM-A42

Benchtop Multi-parameter Water Quality Meter LD-LMPWM-A42

SKU: LD-LMPWM-A42

Benchtop Multi-parameter Water Quality Meter LD-LMPWM-A42 with large backlit LCD (125 × 100 mm) display, measures conductivity; TDS, salinity, resistivity, DO and temperature of the sample. Equipment saves space of the benchtop by combining multiple parameters in one unit. With expanded memory, it stores and recalls up to 500 data sets. Calibration due reminder informs the operator to calibrate the meter regularly. Setup menu allows you to set the number of calibration points, stability criteria, temperature unit, date and time, auto-power off, etc. USB communication interface for data transfer and timed interval readings.

  • Conductivity/ TDS/ Salinity/ Resistivity Features
  • ✓ 1-5 point for calibration with automatically identification of calibration solutions
  • ✓ Can select cell constant (0.1/1/10) and normalization temperature (20°C or 25°C)
  • ✓ TDS conversion factor, seawater, practical salinity measurements mode to meet different water quality measurement requirements
  • ✓ Linear or pure water compensation improves the measurement accuracy
  • DO Features
  • ✓ 1-2 points calibration using the air-sealed water & zero oxygen solution
  • ✓ Manual salinity & barometric pressure compensation for accurate readings
  • Common Features
  • ✓ Auto-read feature to lock and read the endpoint.
  • ✓ Stability indicator to record stable measurement reading.
  • ✓ Reset feature automatically sets all functions to default condition.
  • ✓ Automatic interval readings, sends the measurement data to the computer/printer.
  • ✓ USB interface to transfer stored data.
  • ✓ Allows transferring data to a computer or importing the data to Excel by free software.
MemoryStores up to 500 data sets
OutputUSB communication interface
Weight1.5 kg
DisplayCustom LCD (125 × 100 mm)
TDS Mode
ConnectorBNC, 3.5 mm jack socket
TDS Range0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt
TDS Factor0.1 to 1.0 (default 0.5)
Temperature
Power Supply5V DC power adapter
TDS Accuracy±1% F.S.
Cell ConstantK=0.1 / 1 / 10 / Custom
Salinity Mode
Reset FunctionYes
Salinity Range0.00 to 80.00 ppt / 0.00 to 42.00 psu
TDS Resolution0.01, 0.1, 1, automatic
Resistivity Mode
Compensation ModeYes
Conductivity Mode
Resistivity Range0.00 to 20.00 MΩ
Salinity Accuracy±1% F.S.
Conductivity Range0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Data Hold FunctionManual / Auto-read measurement endpoint
Stability CriteriaLow / High
Salinity Resolution0.01
Other Specifications
Resistivity Accuracy±1% F.S.
Calibration Due Alarm1 to 31 days or off
Conductivity Accuracy±0.5% F.S.
Dissolved Oxygen Mode
Reference Temperature20 / 25 °C
Conductivity Solutions10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS
Dissolved Oxygen Range0.00 to 20.00 mg/L / 0.0 to 200.0 % saturation
Resistivity Resolution0.01
Conductivity Resolution0.001, 0.01, 0.1, 1, automatic
Dimension (L × W × H)210 × 188 × 60 mm
Dissolved Oxygen Accuracy±0.2 mg/L / ±2.0 %
Dissolved Oxygen Resolution0.01 / 0.1, selectable
Temperature Compensation Modemanual or automatic
Temperature Measurement Range0 to 105°C / 32 to 221°F
Temperature Compensation Range0 to 100°C / 32 to 212°F
Conductivity Calibration Points1 to 5 points
Temperature Measurement Accuracy±0.5°C / ±0.9°F
Dissolved Oxygen Calibration Point1 or 2 points
Temperature Measurement Resolution0.1°C/°F
Conductivity Temperature Coefficient0.0 to 10.0% /°C
Dissolved Oxygen Pressure Correction60.0 to 112.5 kPa / 450 to 850 mmHg, manual
Dissolved Oxygen Salinity Correction0.0 to 50.0 g/L, manual
Temperature Measurement Calibration Points1 point

Multi-parameters are used in research testing, agriculture , quality analysis of water, wastewater treatment, other general liquids across water plants, food and beverages production units, laboratories, and institutes.

Similar Products