OurLD-LKV-A13is a double layered, desktop structured Kinematic Viscometer which is designed as per ASTM D2171 / D2171M - 10 “Asphalt Kinematic Viscosity Test (Capillary Viscometer Methodsâ€.
| Bath cubage | 23 L |
| Stirring motor | 6 W, 1200 rpm |
| Sample quantity | 3 piece at the same time |
| Bath temperature | Ambient temperature ~ 180.0 °C |
| Measurement range | 100 ~ 150 °C or 150 ~ 200 °C |
| Relative humidity | ≤ 85 % |
| Ambient temperature | (-10) °C ~ 35 °C |
| Capillary viscometer | 7 pieces of Cannon-Fenske reserve-flow capillary viscometer tubes in a group |
| Temperature controlling accuracy | ± 0.1 °C |
| Mercury thermometer scale division | 0.1 °C |
| Inner capillary viscometer diameters | 1.02, 1.26, 1.48, 1.88, 2.20, 3.10 and 4.00 mm |
It is used in the fields of petroleum & its products industries, research and development department.
Ash Content Tester LACT-A10
✓ Controlled temperature : 775 ±25 ℃
✓ Temperature rising time of empty chamber : ≤ 50 min
Mercaptan Sulfur Tester LMST-A10
✓ Potential measuring range : (0 to ±1999.5) Mv
✓ Intrinsic error : 0.1 % ±0.5 mV
Interfacial Tension Tester LIT-A10
✓ Repeatability 0.3% Resolution 0.1 mN/m Sensitivity 0.1 mN/m Accuracy 0.1 mN/m
Water Soluble Acid and Alkali Tester LAAT-A10
✓ Graduated flask : 100 ml,50 ml
✓ Conical flask : 100ml
Acid Number and Acidity Tester LAAT-A11
✓ Ambient temperature : ≤ 35 °C
✓ Relative humidity : ≤ 85 %