HPLC system LD-LHLC-B11

HPLC system LD-LHLC-B11

SKU: LD-LHLC-B11

This HPLC system delivers high quality performance and reliable results in order to meet the stringent requirements of the customer. To meet GMP/GLP requirements of the instrument, this HPLC system has hardware with VP function which records maintenance information and operations.

  • ✓ High pressure constant-current infusion pump
  • ✓ Even mixing of the injections using mixers
  • ✓ Supreme appearance to meet the customer’s visual requirements
  • ✓ Simple, easy to use operations
  • ✓ UV detector unit consists of dual beam monochromator, light battery, deuterium lamp
Also available in:
High pressure pump : Single pump High pressure pump : Two sets of pump High pressure pump : High pressure gradient pump High pressure pump : High pressure constant-current infusion pump
Weight11 kg
Flow rate≤ 0.5 %
UV detector
Infusion WayMicro volume series dual-piston
Linear range≥ 104
Optical path10 nm
Static noise≤ ± 0.5 × 10-5 A (empty cell, response time, 1 second, 20 ºC)
Air tightnessPressure 5400 Psi, 10 min time, pressure drop < 400 Psi
Dynamic noise≤ ± 0.75 × 10-5 AU (methyl alcohol 1 mL / min, 254 nm, 20 ºC)
Gradient error± 1 % (0 - 100 % water/acetone solvent)
Flow rate range0.001 - 9.999 mL / min (in 0.001 mL / min increments)
Flow cell volume8 µL
Wavelength error≤ 1 nm
Infusion pressure0 - 9999 Psi
High pressure pumpHigh pressure constant-current infusion pump
Limit of detection≤ 1 × 10-9 g / ml (Naphthalene methyl alcohol solvent)
Pressure pulsation< 15 Psi (flow rate 1 mL / min, pressure 600 - 1600 Psi)
Size (W × H × D)260 × 130 × 420 mm
Spectrum band width6 nm
Operation Temperature4 - 40 ºC
Operation temperature4 - 40 ºC
Static baseline Drift≤ 0.5 × 10-4 (empty cell, response time, 1 second, 20 ºC)
Time program functionYes
Dynamic baseline Drift≤ ± 1 × 10-4 AU / h (methyl alcohol 1 mL / min, 254 nm, 20 ºC)
Qualitative repeatabilityRSD 6 ≤ 0.5 %
Quantitative repeatabilityRSD 6 ≤ 0.1 %
Wavelength repeatability error≤ ± 0.1 nm
Stability error for the flow rate≤ 0.2 % RSD

This HPLC has applications in the scientific research studies, detection of residual amount in medicine, food, alcohol and in chemical analysis.