Features

  • UV sterilization and hibernation mode
  • Anti-fog heated observation window
  • Chromatic liquid crystal touch screen interface
  • Built-in waste tray and instrument rack
  • Blade ejector mechanism
  • Smooth handwheel with ergonomic design

Cryostat Microtome LD-LCRY-A21

Cryostat Microtome LD-LCRY-A21 has a section thickness range that can be adjusted from 1 to 100 µm, and trimming thickness up to 600 µm. It combines an efficient freezing system with a semi-automatic microtome for consistent and accurate cryosectioning. This cryostat features a modular control system that supports both automatic and manual defrosting. The built-in waste tray and instrument rack enhance convenience and cleanliness. Our microtome is designed for slicing thin sections of biological tissues at low temperatures.

SKU Code: LD-LCRY-A21

Also Available in below capacities

Power 1400 W at 220 to 240 V 1200 W at 110 to 127 V
Net Weight 120 kg
Refrigerant R4490
Defrost Time 15 minutes
Gross Weight 208 ± 10 kg
Power Supply AC 220 to 240 V 50 Hz AC 220 to 240 V 60 Hz AC 110 to 127 V 60 Hz
Chamber Capacity 20 specimen disc
Packing Dimension 913 × 804 × 1467 mm
Dimension (W×D×H) 780 × 700 × 1280 mm
Maximum Specimen Size 55 × 55 mm
Operating Noise Level Up to 70 dB(A)
Refrigeration Capacity 690 W
Section Thickness Range 1 to 100 µm
Specimen Vertical Stroke 59 mm ± 0.5 mm
Trimming Thickness Range 1 to 600 µm
Fast Forward/Rewind Speed 900 µm/s
Slow Forward/Rewind Speed 300 µm/s
Atmospheric Pressure Range 700 to 1060 hPa
Maximum Operating Altitude Up to 2000 m
Section Thickness Settings 1 μm to 20 μm, in 1 μm increments 20 μm to 50 μm, in 2 μm increments 50 μm to 100 μm, in 5 μm increments
Specimen Horizontal Stroke 25 mm ± 1 mm
Section Thickness Deviation ± 20%
Trimming Thickness Settings 1 μm to 50 μm, in 5 μm increments 50 μm to 100 μm, in 10 μm increments 100 μm to 600 μm, in 50 μm increments
Minimum Reaching Temperature -60°C (±5°C)
Relative Humidity for Storage Up to 60% (non-condensing)
Freezing chamber Defrost Method Manual/ auto
Quick-freezing Shelf Temp Range -40°C to +40°C
Specimen Clamp Adjustable Angle 12° (X & Y axis), 360° (Z axis)
Specimen Holder Freezing Points 18 (mechanical), 2 (semiconductor)
Freezing Chamber Temperature Range -35°C to -40°C
Our Cryostat Microtome LD-LCRY-A21 is used for rapid cryosectioning of tissues in clinical pathology and intraoperative diagnostics. It enables precise preparation of frozen tissue samples for histological, immunohistochemical, and molecular analysis.

Similar Products